When it comes to plant measurements in electroplating baths or dyestuffs, the Metrohm 826 pH mobile with the Metrosensor Unitrode will perfectly do the job!
pH Meter:
Electrode:
- Aquatrode plus – combined LL pH glass electrode with fixed ground sleeve diaphragm and double-junction reference system. Its new type of membrane glass had especially been developed for pH measurement in weakly buffered samples with low or high conductivity
Accessories:
- Cable, 1m (other lengths available)
- Buffer solutions
- pH 4, pH 7 and pH 9 (30x30 mL sachets)
- Electrolyte solution c (KCl) = 3 mol/L, 250mL (for Ag/AgCI reference system)
- Storage solution for pH glass electrodes
Optional
- 826 pH mobile with carrying case and calibration accessories
Your advantages:
- The 826 pH mobile can be purchased together with a case. This way the transport of the meter and the electrode is much easier. The case includes also all accessories you need for calibrating the pH measurement system in the field
- The fixed ground joint diaphragm of the Unitrode stops back diffusion of the sample into the reference chamber of the electrode. The electrolyte solution can not be contaminated with sample
Practical hints:
- pH measurement is temperature dependent. Don't forget to enter the temperature value of your sample solution into the meter. If the temperature changes during the measurement it is recommended to use an Aquatrode Plus with integrated Pt 1000 temperature sensor
- Always store the electrode in the Storage Solution when not in use. This way you'll always have a "ready to go" electrode of optimal performance. Electrode holder and electrolyte solution can be carried quite handy in the accessories case of the 826 pH mobile